A miniaturized hydrogen flame ionization detector based on integrated nozzle assembly and embedded sealing structure | |
Li, Shenghong2; Geng, Xuhui2; Ding, Kun2; Gao, Yan2; Guan, Yafeng1,2 | |
刊名 | TALANTA |
2023-12-01 | |
卷号 | 265页码:7 |
关键词 | Hydrogen flame ionization detector Nozzle with embedded sealing structure Miniaturization Polyimide Reduced gases consumption |
ISSN号 | 0039-9140 |
DOI | 10.1016/j.talanta.2023.124806 |
通讯作者 | Guan, Yafeng(guanyafeng@dicp.ac.cn) |
英文摘要 | A modified miniaturized hydrogen flame ionization detector (m-FID) was developed and evaluated. An integrated nozzle assembly was constructed to solve the gas leakage caused by adhesive crack during repeated highlow temperature processes or vibration. An embedded sealing structure was designed to realize the face sealing, thus improving the sealing stability and reliability of the m-FID. Polyimide was employed as seal and insulation material to ensure the detector can be used at 300 & DEG;C for a long term. The hydrogen and air consumption of the m-FID was 12 mL/min and 110 mL/min, which is about 1/3 of the FID gas consumption of commercial laboratory instruments. The limit of detection (LOD) for n-hexadecane was 3.2 x 10-12 g/s, with a linear response range of nearly 5 orders of magnitude. Finally, it was installed onto an on-site gas chromatograph to detect drug samples with wide boiling point range from room temperature up to 535 & DEG;C. |
资助项目 | Natural Science Foundation of China[YJKYYQ20210038] ; Instrument Developing Project of the Chinese Academy of Sciences ; [22027809] |
WOS研究方向 | Chemistry |
语种 | 英语 |
出版者 | ELSEVIER |
WOS记录号 | WOS:001058592800001 |
资助机构 | Natural Science Foundation of China ; Instrument Developing Project of the Chinese Academy of Sciences |
内容类型 | 期刊论文 |
源URL | [http://ir.idsse.ac.cn/handle/183446/10459] |
专题 | 深海科学研究部 |
通讯作者 | Guan, Yafeng |
作者单位 | 1.Chinese Acad Sci, Inst Deep Sea Sci & Engn, 28 Luhuitou Rd, Sanya 572000, Peoples R China 2.Chinese Acad Sci, Dalian Inst Chem Phys, Dept Instrumentat & Analyt Chem, CAS Key Lab Separat Sci Analyt Chem,Key Lab Deep S, 457 Zhongshan Rd, Dalian 116023, Peoples R China |
推荐引用方式 GB/T 7714 | Li, Shenghong,Geng, Xuhui,Ding, Kun,et al. A miniaturized hydrogen flame ionization detector based on integrated nozzle assembly and embedded sealing structure[J]. TALANTA,2023,265:7. |
APA | Li, Shenghong,Geng, Xuhui,Ding, Kun,Gao, Yan,&Guan, Yafeng.(2023).A miniaturized hydrogen flame ionization detector based on integrated nozzle assembly and embedded sealing structure.TALANTA,265,7. |
MLA | Li, Shenghong,et al."A miniaturized hydrogen flame ionization detector based on integrated nozzle assembly and embedded sealing structure".TALANTA 265(2023):7. |