Parameter Configuration and Data Accuracy Analysis for Real-Time RFI Detection | |
Wang, Yue; Liu, Qi | |
刊名 | Radio Science Letters
![]() |
2020 | |
卷号 | 2 |
ISSN号 | 1024-4530 |
DOI | 10.46620/20-0043 |
产权排序 | 1 |
英文摘要 | Real-time detection of radio frequency interference requires the spectrum analyzer to have a high sweep speed and high data accuracy at the same time. We first analyzed key parameters and the measurement-time calculation method of the spectrum analyzer, then proposed a measurement-parameter configuration method. With this method, the actual measurement time and measurement uncertainty were tested and analyzed using three kinds of spectrum analyzers. The test results showed that the actual measurement time and measurement uncertainty of Instrument C are below 70 ms and 0.031 dB, respectively, from 100 MHz to 13 GHz, which can be better applied to real-time detection of radio frequency interference. |
URL标识 | 查看原文 |
语种 | 英语 |
内容类型 | 期刊论文 |
源URL | [http://ir.xao.ac.cn/handle/45760611-7/4062] ![]() |
专题 | 微波接收机技术实验室 |
作者单位 | Xinjiang Astronomical
Observatory, Chinese Academy of Sciences, 150 Science 1-Street,
Urumqi, Xinjiang 830011, China |
推荐引用方式 GB/T 7714 | Wang, Yue,Liu, Qi. Parameter Configuration and Data Accuracy Analysis for Real-Time RFI Detection[J]. Radio Science Letters,2020,2. |