Recovering laboratory doses using fine-grained quartz from Chinese loess
Du, JH (Du, J. H.)[1]; Kang, SG (Kang, S. G.)[1]; Lu, YC (Lu, Y. C.)[1]; Wintle, AG (Wintle, A. G.)[2,3]; Wang, XL (Wang, X. L.)[1]
刊名Radiation Measurements
2011-10-30
卷号46期号:10页码:1073-1081
关键词Quartz OSL SAR Optical bleaching
DOI10.1016/j.radmeas.2011.07.022
文献子类期刊论文
英文摘要

Here we report the results of dose recovery experiments carried out on fine-grained quartz from a Holocene sample in Chinese loess. Optical bleaching prior to giving a 9.35 Gy dose in the laboratory was carried out using artificial light sources, namely blue light emitting diodes (LEDs) and a solar simulator, followed by more than 24 h storage at room temperature. Stimulating the quartz with blue LEDs at room temperature resulted in overestimation of the recovered dose, whereas using the same stimulation time at temperatures above 150 degrees C resulted in the correct value for the recovered dose. Exposure to the solar simulator for times less than 200 min resulted in overestimation of the dose, whereas progressive underestimation was found for longer bleaching times. When doses of up 5333 Gy were given ahead of a 200 min exposure to the solar simulator, the dose recovered depended upon the magnitude of the previous dose, thus questioning the general application of a simple dose recovery experiment.

语种英语
内容类型期刊论文
源URL[http://ir.ieecas.cn/handle/361006/10271]  
专题地球环境研究所_现代环境研究室
通讯作者Wang, XL (Wang, X. L.)[1]
作者单位1.Institute of Geography and Earth Sciences, Aberystwyth University, Aberystwyth SY23 3DB, UK;
2.McDonald Institute for Archaeological Research, University of Cambridge, Downing Street, Cambridge CB2 3ER, UK
3.SKLLQG, Institute of Earth Environment, Chinese Academy of Sciences, 10 Fenghui S. Road. Hi-Tech Zone, Xi’An 710075, PR China;
推荐引用方式
GB/T 7714
Du, JH ,Kang, SG ,Lu, YC ,et al. Recovering laboratory doses using fine-grained quartz from Chinese loess[J]. Radiation Measurements,2011,46(10):1073-1081.
APA Du, JH ,Kang, SG ,Lu, YC ,Wintle, AG ,&Wang, XL .(2011).Recovering laboratory doses using fine-grained quartz from Chinese loess.Radiation Measurements,46(10),1073-1081.
MLA Du, JH ,et al."Recovering laboratory doses using fine-grained quartz from Chinese loess".Radiation Measurements 46.10(2011):1073-1081.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace