Experimental measurement of the correlation between CT number and heavy ion range | |
Dai, ZY12; He, PB (reprint author), Chinese Acad Sci, Inst Modern Phys, Lanzhou 730000, Peoples R China.; Ye, F12; Liu, XG12; Li, Q12; He, PB12 | |
刊名 | CHINESE PHYSICS C |
2012 | |
卷号 | 36页码:954-957 |
关键词 | Proton Stopping Power Computed-tomography Radiotherapy Calibration |
ISSN号 | 1674-1137 |
DOI | 10.1088/1674-1137/36/10/006 |
英文摘要 | For precision delivery of the Bragg peak of a heavy-ion beam to a target volume in ion beam therapy, it is necessary to know the tissue stopping power. A general approach to solve this problem in ion beam therapy is to convert X-ray CT (computed tomography) numbers into water-equivalent path length (WEPL) coefficients using a CT-WEPL calibration curve for all voxels traversed by the beam. This work aims at establishing a CT-WEPL coefficient calibration curve for the heavy ion therapy project at IMP, so as to compute the range of carbon ion beams in tissues easily according to the patient CT data. Several tissue-equivalent materials were applied to measure their WEPL coefficients using a high-energy carbon ion beam in this work. A CT-WEPL calibration curve was obtained through fitting the measured data, which can be used directly for dose optimization and facilitates the design of patient treatment plans significantly at IMP. |
URL标识 | 查看原文 |
资助项目 | Science and Technology Development Project of Lanzhou City[2008-sr-10] ; Science and Technology Development Project of Lanzhou City[2009-2-12] |
WOS关键词 | PROTON STOPPING POWER ; COMPUTED-TOMOGRAPHY ; RADIOTHERAPY ; CALIBRATION |
WOS研究方向 | Physics |
语种 | 英语 |
出版者 | CHINESE PHYSICAL SOC |
WOS记录号 | WOS:000309900800006 |
资助机构 | HIMAC [B364] ; HIMAC [B364] ; National Institute of Radiological Sciences ; National Institute of Radiological Sciences ; National Basic Research Program of China (973 Program) [2010CB834202] ; National Basic Research Program of China (973 Program) [2010CB834202] ; National Natural Science Foundation of China [10835011, 10675151, 10805064] ; National Natural Science Foundation of China [10835011, 10675151, 10805064] ; Scientific Technology Research Projects of Gansu Province [0702NKDA045, 0806RJYA020] ; Scientific Technology Research Projects of Gansu Province [0702NKDA045, 0806RJYA020] ; Western Talent Program of Chinese Academy of Sciences [0760160XBO, 0860260XBO] ; Western Talent Program of Chinese Academy of Sciences [0760160XBO, 0860260XBO] |
内容类型 | 期刊论文 |
源URL | [http://ir.impcas.ac.cn/handle/113462/18585] |
专题 | 近代物理研究所_医学物理研究室 近代物理研究所_生物物理研究室 |
通讯作者 | He, PB (reprint author), Chinese Acad Sci, Inst Modern Phys, Lanzhou 730000, Peoples R China. |
作者单位 | 1.Li Qiang 2.Li Qiang 3.Huang Qi-Yan] Chinese Acad Sci, Grad Univ, Beijing 100049, Peoples R China 4.[Dai Zhong-Ying 5.Xiao Guo-Qing] Chinese Acad Sci, Inst Modern Phys, Lanzhou 730000, Peoples R China 6.Huang Qi-Yan 7.Jin Xiao-Dong 8.Liu Xin-Guo 9.Jin Xiao-Dong 10.Liu Xin-Guo |
推荐引用方式 GB/T 7714 | Dai, ZY,He, PB ,Ye, F,et al. Experimental measurement of the correlation between CT number and heavy ion range[J]. CHINESE PHYSICS C,2012,36:954-957. |
APA | Dai, ZY,He, PB ,Ye, F,Liu, XG,Li, Q,&He, PB.(2012).Experimental measurement of the correlation between CT number and heavy ion range.CHINESE PHYSICS C,36,954-957. |
MLA | Dai, ZY,et al."Experimental measurement of the correlation between CT number and heavy ion range".CHINESE PHYSICS C 36(2012):954-957. |
个性服务 |