CORC  > 北京大学  > 工学院
Evaluation of a MEMS-Based Dual Metal-Layer Thin-Film Microelectrode Array for Suprachoroidal Electrical Stimulation
Sui, Xiaohong ; Sun, Jingjing ; Li, Liming ; Zhou, Chuanqing ; Luo, Xuejiao ; Xia, Niansheng ; Yan, Yan ; Chen, Yao ; Ren, Qiushi ; Chai, Xinyu
刊名ieee transactions on neural systems and rehabilitation engineering
2013
关键词Electrically evoked potentials microelectrode polyimide film suprachoroidal implantation visual prosthesis TRANSRETINAL STIMULATION RABBIT EYES MULTICHANNEL ELECTRODE DISSOLUTION PROSTHESIS RETINA
DOI10.1109/TNSRE.2012.2188042
英文摘要A double metal-layer thin-film platinum microelectrode array was fabricated for implantation between sclera and choroid based on MEMS processing techniques and photosensitive polyimide material. The array was composed of 60 stimulating sites (6 x 10) and four selectable returning electrodes. The diameter of each stimulating electrode was 350 mu m with a center-to-center spacing of 750 mu m. The transient voltage responses of the electrode to current pulse stimulation indicated a charge-injection capacity greater than 52.1 mu C/cm(2) . Acute in vivo animal experiments showed that the implicit time of electrically evoked potentials (EEPs) was 17.09 +/- 1.45 ms at a threshold current of 25.55 +/- 5.43 mu A for a full-row of simultaneously stimulated electrodes (i.e., current applied simultaneously to each of the 10 electrodes). Individual electrode stimulation threshold was 48.57 +/- 6.90 mu A. The corresponding threshold charge densities were 13.28 +/- 2.82 mu C/cm(2) and 25.24 +/- 3.59 mu C/cm(2) , respectively. The spatial spread of the maximally recorded P1 response in the EEPs indicated a correspondence between the retinal stimulation site and the focal response location in the cortex. This method of array fabrication is suitable for acute suprachoroidal stimulation, and has a potential use for the fabrication of a visual prosthesis.; http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000321737000002&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=8e1609b174ce4e31116a60747a720701 ; Engineering, Biomedical; Rehabilitation; SCI(E); EI; 3; ARTICLE; 4; 524-531; 21
语种英语
内容类型期刊论文
源URL[http://ir.pku.edu.cn/handle/20.500.11897/222158]  
专题工学院
推荐引用方式
GB/T 7714
Sui, Xiaohong,Sun, Jingjing,Li, Liming,et al. Evaluation of a MEMS-Based Dual Metal-Layer Thin-Film Microelectrode Array for Suprachoroidal Electrical Stimulation[J]. ieee transactions on neural systems and rehabilitation engineering,2013.
APA Sui, Xiaohong.,Sun, Jingjing.,Li, Liming.,Zhou, Chuanqing.,Luo, Xuejiao.,...&Chai, Xinyu.(2013).Evaluation of a MEMS-Based Dual Metal-Layer Thin-Film Microelectrode Array for Suprachoroidal Electrical Stimulation.ieee transactions on neural systems and rehabilitation engineering.
MLA Sui, Xiaohong,et al."Evaluation of a MEMS-Based Dual Metal-Layer Thin-Film Microelectrode Array for Suprachoroidal Electrical Stimulation".ieee transactions on neural systems and rehabilitation engineering (2013).
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace