Calibration of clock scanning microwave interferometric radiometer | |
Zhang, Cheng; Liu, Hao; Wu, Ji | |
2018 | |
会议日期 | March 27, 2018 - March 30, 2018 |
会议地点 | Cambridge, MA, United states |
页码 | 121-124 |
英文摘要 | Clock Scanning Microwave Interferometric Radiometer (CS-MIR) is a new concept of synthetic aperture radiometer. It has the advantages of simple array structure, easy deployment and resultingly being able to achieve high spatial resolution. Recently a ground-based prototype of CS-MIR working at L-band has been developed. This paper presents the calibration method of this instrument. Compared with the SMOS/MIRAS system, the calibration of the CS-MIR is much simple. The simplicity is mainly lies in three aspects: It does not need quadrature error correction because of the digital I/Q demodulation, not need PMS subsystem for visibility demoralization and not need correlated noise injection for phase calibration. © 2018 IEEE. |
会议录 | 15th Specialist Meeting on Microwave Radiometry and Remote Sensing of the Environment, MicroRad 2018 - Proceedings |
语种 | 英语 |
ISBN号 | 9781538650158 |
内容类型 | 会议论文 |
源URL | [http://ir.nssc.ac.cn/handle/122/6407] |
专题 | 国家空间科学中心_微波遥感部 |
推荐引用方式 GB/T 7714 | Zhang, Cheng,Liu, Hao,Wu, Ji. Calibration of clock scanning microwave interferometric radiometer[C]. 见:. Cambridge, MA, United states. March 27, 2018 - March 30, 2018. |