Determination of Metal Elements in PM2. 5 by ICP-OES with Microwave Digestion
Zhang Liu-yi1; Fu Chuan1; Yang Fu-mo1,2; Yang Ji-dong1; Huang Yi-min1; Zhang Qiang1; Wu Bing-yu1
刊名SPECTROSCOPY AND SPECTRAL ANALYSIS
2014-11-01
卷号34期号:11页码:3109-3112
关键词Microwave digestion ICP-OES Fine particulate matter Detection limit Recovery
ISSN号1000-0593
DOI10.3964/j.issn.1000-0593(2014)11-3109-04
通讯作者Fu, C (reprint author), Chongqing Three Gorges Univ, Key Lab Water Environm Evolut & Pollut Control Th, Chongqing 404100, Peoples R China.
英文摘要In the present work, a method was developed for determining lead, zinc, copper, cadmium, znd chromium in PM2. 5 by inductively coupled plasma-optical emission spectrometry (ICP-OES) analysis with microwave digestion and glass fibre filter collection of samples. The microwave digestion systems were investigated and the experimental conditions were optimized. The results show that (1) HNO3-H2O2 digestion system is more stable and complete than HNO3-HCl and HNO3-H2S4 digestion systems; (2) The most sensitive emission wave length of lead, zinc, copper, cadmium, and chromium are 220. 353, 213. 857, 327. 393, 228. 802, and 267. 716 nm, respectively; (3) The highest signal-to-noise ratios were observed under the conditions; RF power of 1 300 W, peristaltic pump flow rate of 1. 5 mL . min(-1), cooling gas flow rate of 15 L . min(-1), and carrier gas flow rate of 0. 8 L . min(-1). In addition, the detection limit for these elements ranged between 2. 02 X 10(-3) and 8. 20 X 10(-3) mu g " mL(-1), the relative standard deviations (RSD, n=6) for the samples were in the range of 1. 86%similar to 2. 82%, and the recovery for the elements determined was from 91. 6% to 103. 7%. The proposed method was used for determination of the above five elements in atmospheric fine particulate matter at Wanzhou Monitoring Site of Chongqing Institute of Green and Intelligent Technology. The results revealed that the atmospheric fine particulate matter at this monitoring site was not polluted by cadmium and chromium, lead was at the level of potential contamination, while zinc and copper were at the level of slight pollution.
WOS研究方向Spectroscopy
语种英语
出版者OFFICE SPECTROSCOPY & SPECTRAL ANALYSIS
WOS记录号WOS:000344447700045
内容类型期刊论文
源URL[http://119.78.100.138/handle/2HOD01W0/1131]  
专题大气环境研究中心
通讯作者Fu Chuan
作者单位1.Chongqing Three Gorges Univ, Key Lab Water Environm Evolut & Pollut Control Th, Chongqing 404100, Peoples R China
2.Chinese Acad Sci, Chongqing Inst Green & Intelligent Technol, Key Lab Reservoir Aquat Environm, Chongqing 400714, Peoples R China
推荐引用方式
GB/T 7714
Zhang Liu-yi,Fu Chuan,Yang Fu-mo,et al. Determination of Metal Elements in PM2. 5 by ICP-OES with Microwave Digestion[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS,2014,34(11):3109-3112.
APA Zhang Liu-yi.,Fu Chuan.,Yang Fu-mo.,Yang Ji-dong.,Huang Yi-min.,...&Wu Bing-yu.(2014).Determination of Metal Elements in PM2. 5 by ICP-OES with Microwave Digestion.SPECTROSCOPY AND SPECTRAL ANALYSIS,34(11),3109-3112.
MLA Zhang Liu-yi,et al."Determination of Metal Elements in PM2. 5 by ICP-OES with Microwave Digestion".SPECTROSCOPY AND SPECTRAL ANALYSIS 34.11(2014):3109-3112.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace