The Method of Linear Inflation Control in Ambulatory Blood Pressure Measurement at Finger
Pandeng Zhang; Jie Zhang; Quanli Qiu; Ye Chen; Jia Liu
2016
会议名称RCAR2016
会议地点吴哥窟,柬埔寨
英文摘要In 24-hour Non-invasive Blood Pressure (NIBP) measurement, most of the automatic devices use oscillometric method. Accurate measurement of BP at finger using an Oscillometric Waveform (OMW) remains a challenge, particularly when the waveform is collected in inflation, because of the bad signal-to-noise ratio. In the present work, we aimed at measuring Ambulatory Blood Pressure by extracting OMW signal at finger with a new method of linear inflation control and digital signal processing. With the PID control of the pump and real time signal processing, we implemented the blood pressure measurement system and new algorithm. For the validation of our suggested method, we compare the effect of our method and traditional method. Finally, the method is validated in a pilot study against Welch Allyn ProBP 3400 BP monitors.
收录类别其他
语种英语
内容类型会议论文
源URL[http://ir.siat.ac.cn:8080/handle/172644/10308]  
专题深圳先进技术研究院_数字所
作者单位2016
推荐引用方式
GB/T 7714
Pandeng Zhang,Jie Zhang,Quanli Qiu,et al. The Method of Linear Inflation Control in Ambulatory Blood Pressure Measurement at Finger[C]. 见:RCAR2016. 吴哥窟,柬埔寨.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace