A study of time over threshold (TOT) technique for plastic scintillator counter
Wu JJ(吴金杰); Heng YK(衡月昆); Sun ZJ(孙志嘉); Wu C(吴冲); Yang GA(杨桂安); Jiang CH(姜春华); Wu, JJ; Heng, YK; Sun, ZJ; Wu, C
刊名CHINESE PHYSICS C
2008
卷号32期号:3页码:#REF!
关键词scintillator counter time over threshold (TOT) charge measurement relative resolution
通讯作者Wu, JJ (reprint author), CAS, Inst High Energy Phys, Beijing 100049, Peoples R China.
英文摘要A new charge measurement method, time over threshold (TOT), has been used in some gas detectors lately. Here TOT is studied for TOF system, made of plastic scintillator counter, which can simplify the electronics of the system. The signal characteristics are measured and analyzed with a high quality oscilloscope, including noise, pedestal, signal amplitude, total charge, rise time and the correlation between them. The TOT and charge are related and can be fitted by some empirical formula. The charge measurement resolution by TOT is given and this will help the design of TOF electronics.
学科主题Physics
类目[WOS]Physics, Nuclear ; Physics, Particles & Fields
收录类别SCI
WOS记录号WOS:000253748600005
公开日期2016-05-03
内容类型期刊论文
源URL[http://ir.ihep.ac.cn/handle/311005/226783]  
专题高能物理研究所_实验物理中心
推荐引用方式
GB/T 7714
Wu JJ,Heng YK,Sun ZJ,et al. A study of time over threshold (TOT) technique for plastic scintillator counter[J]. CHINESE PHYSICS C,2008,32(3):#REF!.
APA 吴金杰.,衡月昆.,孙志嘉.,吴冲.,杨桂安.,...&Jiang, CH.(2008).A study of time over threshold (TOT) technique for plastic scintillator counter.CHINESE PHYSICS C,32(3),#REF!.
MLA 吴金杰,et al."A study of time over threshold (TOT) technique for plastic scintillator counter".CHINESE PHYSICS C 32.3(2008):#REF!.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace